5 data output, 6 operational checkout, 7 counts – Fluke Biomedical 1060AM User Manual

Page 20

Advertising
background image

Victoreen 1060AM
Operators Manual

3-4

7.

Open

8.

ON = Terminal vs Computer communication format (Reserved)

3.5 Data Output

The standard communication is a block of 8-bit ASCII characters. Standard network protocol is defined in
Section 3.1. Terminal data protocol and commands are defined in Section 3.2.

3.6 Operational Checkout

The LEDs indicate normal operation and are defined in Section 4.1.10. A visual review of their status will
provide a quick check of operability of the Model 1060.

3.7 Counts

3.7.1 Counts per second

The number of counts from the detector is accumulated for 0.1 seconds. Ten of these intervals are
summed to arrive at a Counts Per Second (CPS) total.

3.7.2 Counts per minute

Each Counts Per Second (CPS) value is stored in a buffer location. Sixty CPS values are summed to
arrive at a Counts Per Minute (CPM) total.

3.7.3 Counts per hour

Each Counts Per Minute (CPM) value is stored in a buffer location. Sixty CPM values are summed to
arrive at a Counts Per Hour (CPH) total.

3.7.4 Counts in mR/Hr

Counts in mR/Hr are provided using CPM to mR/Hr conversion factor.

3.7.5 Counts in Sieverts

Counts in Sieverts are provided using mR/Hr to Sieverts conversion factor.

3.7.6 Counts per Correction Factor

Counts per correction factor are provided using correction factors 0 through 9.