Victoreen 1060AM
Operators Manual

1-2

NEMA 4X Enclosure


Model No. Description

1060AM-NM-ER

Environmental range - 1 μR to 1000 (μR/hr (SI units)

1060AM-NM-LR

Low range - 0.01 mR to 1000 mR/hr (SI units)

1060AM-NM-MR

Medium range - 0.1 mR to 10,000 mR/hr (SI units)

1060AM-NM-HR

High range - 1 mR to 100,000 mR/hr (SI units)

1060MHV-NM-HR

Area Monitor with external MHV probe connection

1.2 Specifications

1.2.1 Digital Specifications

On-board Computer

Intel 80C196, 16 Bit, 16 MHz, Embedded CMOS Microcontroller

1 Serial Ports

3 asynchronous modes

Up to 187.5 K Baud @ 12 MHz

1 Buffer for input/output

HSI high-speed input capture

Interrupt on either rising, falling or a transition

2

Timers

Timer 1, Free running, based on XTAL

Timer 2, Incremented by T2CLK or HSI.1

512 Bytes of RAM

8 Channel, 10 Bit, ADC w/ Sample and Hold Available at PCB connector

for external inputs. Two channels are dedicated to jam detection
and detector temperature sensing.

Extended temperature range -25°C to +85°C

Memory

PROM

OTP (One-Time-Programmable)

16K X 16, (256K) Program/Data

48K X 16, (768K) Paged

RAM

Static RAM, 64K X 16 (1Mb)

EEPROM Serial IK X 16, for conversion factors, setpoints, historical data (stored

upon loss of communications only), etc.

Serial Communications

Dual UART: Two RS-485 and one RS-232

The RS-485 is the primary communications link to data acquisition system. The
RS-232 is a local diagnostics, local data logging, and set-up port.

Status Indicators

8 LEDs indicating operating status (i.e., communications in progress, normal
operation, high voltage ON/OFF)

Switches

8 DIP switches used to initiate various local functions (i.e., high voltage, on/off,

etc.)

Digital Outputs

16 latched TTL outputs