OWON HDS1021M Handheld DSO & DMM 2-Safety Information

adapter appointed by the manufacturer and subject to approval of being used in the user’s
country.

Warning:

The two channels of the oscilloscope are non-isolated electrically. The channels should adopt
common basis during measuring. To prevent short circuits, the 2 probe ground must not be
connected to 2 different non-isolated DC level.

Warning:

The channels should adopt common basis during measuring. To prevent short circuits, the 2
probe ground must not be connected to 2 different non-isolated DC level. The diagram of the
oscilloscope ground wire connection:

It is not allowed to measure AC power when charging the oscilloscope with adapter.

Warning:

To avoid fire or electrical shock if a test tool input is connected to more 42V peak (30Vrms)
or on circuits of more than 4800VA:

●

Use only insulated voltage probes, test leads and adapter supplied with the test tool, or
indicated by OWON as suitable for the Oscilloscope & Multimeter.

●

Before use, inspect voltage probes, test leads and accessories for mechanical damage and
replace when damaged.

●

Remove all probes, test leads and accessories that are not in use.

●

Always connect the power adapter first to the AC outlet before connecting it to the
Oscilloscope & Multimeter.

●

Do not apply voltages that differ more than 400 V from earth ground to any input when
measuring in a CAT
Ⅱ environment.

●

Do not apply voltages that differ more than 400 V from each other to the isolated inputs
when measuring in a CAT
Ⅱ environment.

●

Do not apply input voltages above the rating of the instrument Use caution when using
1:1 test leads because the probe tip voltage will be directly transmitted to the
Oscilloscope & Multimeter.

●

Do not use exposed metal BNC or banana plug connectors.

●

Do not insert metal objects into connectors.

●

Always use the Oscilloscope & Multimeter only in the manner specified.