READ THE RESULTS

The result is an estimate of the subject’s BAC. A breath alcohol tester

cannot determine the exact BAC in the blood of the subject. Correlation

between BrAC and BAC depends on many factors, such as temperature

and health conditions. Only use this device to screen for the presence

of alcohol.
Many people consider a 0.02% or greater a positive test for alcohol in

“zero tolerance” screening. Most states have 0.08% as the legal BAC

limit for tasks such as operating a motor vehicle. Some states prohibit

driving commercial vehicles or any vessel with a 0.04% BAC or higher.

Many states prosecute individuals for driving under the influence at any

level.

ERROR MESSAGES

Error Message

Cause and Course of Action

The unit displays “WARN” and plays alarm tones when

the result is higher than 0.05% BAC.

The unit did not detect an even alcohol concentration in

the breath sample. This can occur if the user varies the

blowing force when performing a test or if the exhaust port

is obstructed. Repeat operation steps 1 – 6 to perform

another test.
The unit displays “Hot” and plays alarm tones when it

has detected a very high alcohol concentration (above

0.40% BAC).

Battery power is low. Install a new 9V alkaline battery.

Note: Any of these errors will turn the unit’s backlight red.