Ion Science TVOC User Manual

Page 25

Advertising
background image

TVOC MANUAL

Ion Science Ltd

Page 25 of 32

Unrivalled Detection. www.ionscience.com

System recommendations


For a gas sample system we have the following recommendations:

1. Where possible use the Flow Adaptor (as described above) that is sold by ISL. This has an integral

o-ring to seal the sensor housing and ports for connection of inlet and out let sample tubes. Please refer
to Spare parts section for part number.

2. The pressure difference of a pumped system relative to atmosphere should be minimised to avoid the

effects of gas law.

3. The maximum pressure that can be applied to the TVOC sensor housing is 300 mbar. However this is

not a recommended working pressure. Ideally working pressure should be +/- 30 mBar relative to
ambient pressure.

4. In line flow restrictions must be minimized. Flow restrictions cause differential pressures which can

directly affect the TVOC ready. If flow restrictions are unavoidable the flow rate should be lowered to
minimize the pressure effects, this however will increase the response time.

5. A flow between 250 and 500 ml/min is recommended for calibration. This will ensure a full response to

applied gas in a sensible time.

6. The working flow should be very similar to that use to calibrate the instrument or errors in output will

occur (see point 2).

7. The response time of the system is defined by the sensor response rate and sample flow rate will in

combination with the tube length and diameter as well as any dead volumes.