Technical specifications – Ion Science TVOC User Manual

Page 30

Advertising
background image

TVOC MANUAL

Ion Science Ltd

Page 30 of 32

Unrivalled Detection. www.ionscience.com

Technical specifications

PID Sensor

With fence electrode technology

Lamp type

10.6 eV (Krypton)


TVOC enclosure ingress protection

IP65

Sensor ingress protection

IP54


Range

0

– 10 ppm / 0 – 22.8 mg/m

3

(0.01 resolution)

0

– 100 ppm / 0 – 228 mg/m

3

(0.1 resolution)

0- 1,000 ppm / 0

– 2280 mg/m

3

(1.0 resolution)


Non IS applications

Input power

5

– 28 Vdc. 65 mA (0.5mm

2

to 2.5mm

2

C.S.A.)

4

– 20 mA power

8

– 35 Vdc. 22 mA (0.5mm

2

to 2.5mm

2

C.S.A.)

4-20mA loop must be externally powered


ATEX approval

II 2G EEx ia IIC T4 (-

20°C ≤Ta ≤+50°C)

IECEx approval

Ex ia IIC T4 (-20°C Ta +50°C)

Certificate number

Baseefa05ATEX0277

IS applications

Entry parameters

Input power

Ui = 18V

Ii = 800mA

Pi = 1.2W

Ci = 0

F

Li = 0mH

4-20mA power

Ui = 30V

Ii = 200mA

Pi = 1.2W

Ci = 0

F

Li = 0mH

For information only please see ATEX certificate before installation

Zener Barriers:

Required Ask your installation engineers for installation/application advice.



(Non sparking) approval

II 3G Ex nA IIC T4 (-40

o

C < Ta < +50

o

C) (nA for zone 2 only)



Dimensions

Weight

Height

188 mm

(7.40

”)

Instrument

1.3kg

(2.9 lb)

Width

126 mm

(4.96

”)

Packed

1.47kg

(3.2lb)

Depth

78 mm

(3.07

”)


Display

7 Segment, 4 Digit LCD. 4 Colour LED’s

Response

Sensor T90 < 5sec
TVOC output update: Variable 1 second to 10 minutes

Accuracy

0 to 100 ppm: +/- 1 ppm (whichever is greater)

100 to 1000 ppm: +/- 10% or +/- 10 ppm (whichever is greater)

Linearity

0

– 1000 ppm >75%

Calibration

Magnetically accessed

ZERO = Carbon canister
SPAN = 100 ppm Isobutylene +/- 10%

Temperature

Operation: -20

o

C to +50

o

C (-4

o

F to 122

o

F)

Humidity

0

– 95% RH (non-condensing)


EMC

Screened cables are required to archive the light industrial immunity

levels.

NB: All specifications are against an isobutylene calibration at 20 °C, 90% RH and up to 100 ppm.