Casio E-CON Version 1.20 User Manual

Page 13

Advertising
background image

(3) Real-time Sampling Channel (Use CH)

• 1(CH1) ........ Channel 1

• 2(CH2) ........ Channel 2

• 3(CH3) ........ Channel 3

• 4(SONIC) .... Sonic channel

• Note that the above options appear only when real-time sampling is turned on (by

pressing 1(YES) for the Real-Time item).

(4) Filter Settings (Filter)

• 1(None) ....... No setting

• 2(S-G) ......... S-G Smoothing

b(5-p): 5-point

c(9-p): 9-point

d(17-p): 17-point

e(25-p): 25-point

• 3(MED) ....... Median Filter

b(3-p): 3-point

c(5-p): 5-point

• Note that the above options appear only when real-time sampling is turned off (by

pressing 2(NO) for the Real-Time item).

3. After all the settings are the way you want, press w to return to the Advanced Setup

menu.

u To configure a custom probe

You can use the procedures in this section to configure a custom probe*

1

for use with the

EA-100.

Creating a New Custom Probe Configuration

To configure a custom probe, you must input values for the constants of the fixed linear
transformation formula (

ax

+

b

). The required constants are slope (

a

) and intercept (

b

).

x

in the above expression (

ax

+

b

) is the sampled voltage value (sampling range: 0 to 5

volts).
You can use either of the two following procedures to create a new custom probe
configuration while creating an EA-100 setup using Advanced Setup.

2-11

EA-100 Setup

*

1

The term “custom probe” means any sensor
other than the CASIO or VERNIER sensors
specified as standard for the E-CON Mode.

Advertising