Heliox / Oxygen Blender Service Manual

Page 11

LOW FLOW

Figure J

11. Thread new primary outlet assembly [37] to Manifold Block torque to 10 ft-lbs

(13.6 Nm).



Figure K

12. Start threads of new alarm assembly [11] by hand, tighten with Retaining Ring Pliers, ensure not to

bend reed.

13. Attach bottom cover [6] using four flat head screws [2].
14. Install new Oxygen inlet assembly [5] torque to 10 ft-lbs (13.6 Nm). Oxygen inlet assembly has left

handed threads.

15.

Install new Heliox inlet assembly [4], torque to 10 ft-lbs (13.6 Nm).

Figure L

16. Lubricate proportioning valve bore with Krytox

GPL 106.

17. Align the (3) holes on the proportioning valve

assembly [16] equal distance between the (2)
plugs and push in. *Reference drawing below.

18. Replace washer [19] and phillips head screw

[18].

19. Place new resistance ring [17] in its place on the

proportioning valve assembly.
*Proportional valve assembly inserted, with
the “L” stamped on back for Low Flow Model.