Step 3: low flow assembly, High flow, Tools required – Precision Medical HeliO2 Blenders User Manual

Page 21: Figure h, Figure i

Advertising
background image

Heliox / Oxygen Blender Service Manual

Page 19

HIGH FLOW

Step 3: LOW FLOW ASSEMBLY

Tools Required

Lint Free Swab (optional)
Pointed instrument for removing O-rings
Krytox GPL 106 or equivalent
Oxygen safe lubricant
#2 Phillips Screwdriver
1/2 in. Open End Torque Wrench
(Torque wrench(s) capable of 60 in-lbs and 10 ft-lbs)
11/32 in. Nut Driver
5/32 in. Long Hex Key
Small Retaining Ring Pliers

Figure H

1. Position manifold block so that the mounting bracket is facing up. The large

holes opposite the mounting locations of the diaphragm housing assemblies
are facing you.

2. Cover the alarm assembly thread location with thumb. Drop spring [25] then

ball [24] into alarm bypass orifice.

3. Place alarm bypass body [23] onto long shaft of 5/32 hex key. Guide the alarm

bypass body, threads first into the alarm bypass orifice, and screw into cavity.
Tighten until alarm bypass body is flush with bottom of center hole. Tighten
enough to clear center threads ONLY.

4. Replace O-ring [32] on plug [22]. Insert plug and tighten.


Figure I

5. Lubricate alarm bypass assembly bore on manifold block with Krytox

GPL 106.

6. Using retaining ring pliers install retaining ring [12] in one side of alarm

assembly bore.

7. Insert alarm assembly [13] into bore of manifold block.
8. Install remaining retaining ring.
9. Thread auxiliary outlet assembly [10] to manifold block, torque to 10 ft-

lbs (13.6 Nm).

Advertising