Heliox / Oxygen Blender Service Manual

Page 7

SECTION 3: MAINTENANCE PROCEDURES, REPAIR AND CALIBRATION

I. LOW FLOW (PM5400 Model)

Step 1: DISASSEMBLY

Tools Required

#2 Phillips Screwdriver

11/32 in. Nut Driver

#3 Phillips Screwdriver

5/32 in. Long Hex Key

1/2 in. Open End Wrench

Small Retaining Ring Pliers

Figure A

1. Rotate dial [1] to the 60 graduation.
2. Remove the two flat head screws [2] on each side of the

top cover [3].

3. Remove top cover by pulling upwards.

The cover will not come off unless the dial is at the 60
graduation







Figure B

4. Use a ½ in. open end wrench to unscrew and remove the

Heliox [5] and Oxygen [4] inlet assemblies from bottom of the
Blender.
Oxygen inlet has left hand threads.

5. If manifold outlet assembly is present, unscrew the stem using

a 5/32 hex key and holding the manifold outlet assembly to
the bottom of the Blender.

6. Remove the four flat head screws [2] from bottom cover [6].
7. Remove bottom cover.