Perform a Blower Door test on a room and

record the EqLA at 50 Pa.

Install cardboard in upper part of doorway with a

20 x 20 inch hole cut in it.

Perform a second Blower Door test on the

building, record the EqLA at 50 Pa.

Subtract the first result from the second result

and the value should be 400 sq. in. (+/-10%).

Speed control is handled automatically with FanTestic
software, for complete automation.

Control fan speed with software

Field check system monthly

Alternatively, use a Retrotec Calibration Verifier.

Page 4 of 4

Field check gauge weekly

Check for gauge accuracy, blocked, leaking, or pinched

tubes weekly, and anytime results are in question.

Check the gauge with the tubes in the Umbilical.

Set [Time Average] to 5 seconds in [Settings].

Tap [Channel B] and select“Pressure: Pa”.

Connect the red tube between the red and

yellow ports.

Remove red tubes and repeat with the yellow tube

between the blue and green ports.

If readings on Channel A and Channel B are within

2% and don’t drop noticeably in 5 seconds, the

gauge is functioning correctly.

Check gauge and tubes

regularly to eliminate

the most common

source of errors.