2 quick boot, Quick boot -18, Ar. 3.2.17.2) – KEPCO KLR Series User Manual User Manual

Page 58: R. 3.2.17.2), 2a) or

Advertising
background image

3-18

KLR091313

2. To select mode rotate either VOLTAGE or CURRENT control until status display reads FLT

(factory), MAN (manual), or LIST. Then press DC OUTPUT switch to save selection.

a. If FLT is selected, rotate VOLTAGE or CURRENT control to select NORM, V->C, or

C->V. in the status display.

(1) To select NORM, tap the DC OUTPUT switch to save selection; unit returns to RELY

menu.

(2) To select V->C or C->V, first set the delay. While V->C or C->V is in the status dis-

play, the DC AMPERES display shows the delay in seconds, from 000.5 to 128.0:
NNND, where NNN is an integer from 0 to 128 and D will be either .0 or .5 (000.0 is
not permitted). The delay determines the length of time that the unit must be operat-
ing in the new mode before energizing the relay.

The factory default delay time is set to 1.0 seconds to prevent false tripping of the
relay during application of output power to low impedance or highly reactive loads. If
false tripping occurs, this delay may be increased as needed.

To change the delay, press in and rotate either the VOLTAGE or CURRENT control;
release the control when the DC AMPERES display shows the desired delay. Tap
the DC OUTPUT switch to save selection; unit returns to RELY menu.

NOTE:

Once the relay turns on in V->C or C->V mode, the relay will go off again when the
unit transitions back to the original mode, or when the output is turned off.

b. To select MAN, tap the DC OUTPUT switch to save selection. The DC VOLTS display

flashes the present relay state: ON or OFF. To change the state of the relay rotate either
VOLTAGE or CURRENT control until the desired relay state is displayed, then tap the
DC OUTPUT switch to immediately set the relay to the selected state and return to the
RELY menu.

c. To select LIST, tap the DC OUTPUT switch to save selection; unit returns to the RELY

menu

3. Rotate either VOLTAGE or CURRENT control until status display shows RTN and press DC

OUTPUT switch to exit UTIL menu, then rotate either VOLTAGE or CURRENT control until
status display shows EXIT and press DC OUTPUT switch to exit FUNCTION control.

NOTE: If a SYSTem:SECurity:IMMediate command is issued, the above relay control settings

are reset to the factory default: FLT - NORM.

3.2.17.2 QUICK BOOT

Enter UTIL menu (see PAR. 3.2.17, steps 1 through 3) and rotate either VOLTAGE or CUR-
RENT
control until status display reads QkBT - Quick Boot. If Quick boot is ON, then the next
time the unit is powered up it will go directly to setpoint mode (status display reads SET, see
PAR. 3.2.10) without displaying MODL, LMT, and PROT settings.

• The DC VOLTS display indicates whether quick boot is ON or OFF.

1. Tap the DC OUTPUT switch to change the setting. The DC VOLTS display begins to blink.