15xp-b electrical specifications – Amprobe 5XP-A 15XP-A 35XP-A User Manual

Page 22

Advertising
background image

15

16

BATTERY TEST
Ranges:
1.5 V, 9 V
Accuracy: ± (3.5 % rdg + 2 dgts)

OVERLOAD PROTECTION
Voltage, Resistance, Diode, Continuity:
1000 VDC or 750 VAC rms
200 mV Range: 1000 VDC / 750 VAC rms (3 minutes)
Current:
0.315 A / 1000 V fast blow ceramic fuse 6.3 × 32 mm
mA JACK: Input warning detects wrong switch/input jack
configuration

AUXILIARY FEATURES
DATA HOLD:
Freeze the latest reading on the display.
MIN/MAX: Record the maximum and minimum reading in a
measurement.

15XP-B Electrical Specifications

(at 23°C ± 5°C, < 75 % R.H.)

DC VOLTS
Ranges:
200 mV, 2 V, 20 V, 200 V, 1000 V
Accuracy: ±(0.5 % rdg + 1 dgt)

AC VOLTS

(45 Hz to 500 Hz)

Ranges: 200 mV, 2 V, 20 V, 200 V, 750 V
Accuracy: ±(1.5 % rdg + 5 dgts) 45 Hz to 100 Hz on 200 mV
range: ±(1.5 % rdg + 5 dgts) on 2 V to 750 V ranges

DC CURRENT
Ranges:
200 µA, 2000 µA, 20 mA, 200 mA, 2 A
Accuracy: ±(1.0 % rdg + 2 dgts) on 200 µA to 200 mA ranges:
±(2.0 % rdg + 3 dgts) on 2 A range

AC CURRENT

(45 Hz to 500 Hz)

Ranges: 200 µA, 2000 µA, 20 mA, 200 mA, 2 A
Accuracy: ±(1.5 % rdg + 5 dgts) on 200 µA to 200 mA ranges:
±(2.5 % rdg + 5 dgts) on 2 A range

RESISTANCE
Ranges:
200 , 2 k, 20 k, 200 k, 2 M, 20 M, 2000 M

Advertising