Once conversion has been completed, a “total reset” of the CPU-PCB must be
done.

Record all end-user (custom) programs before undertaking a Total Reset.

■

Secure all panels and covers.

Total reset

■

Turn on unit.

■

Press and hold IQT Sensor key, IQT key, and PROG/START
simultaneously and turn unit off.

■

Release all keys. Wait at least 4 seconds and turn unit on.

NOTE: With total reset procedure all custom programs are deleted. After a total
reset, 12:00 or 13:00 will be indicated on the timer display. Real time must be
reset. To set clock to actual time, see page 50.

Humidity calibration

■

The interior cabinet must be clean. If it is not, wipe clean.

■

The interior cabinet temperature must be below 122°F (50°C) and the
cabinet must be dry.

■

Turn on unit.

■

Open cabinet door.

■

Select any cooking mode.

■

Access diagnostic program by pressing TEMPERATURE key, IQT
Sensor key, and IQT key simultaneously until the word “VERSION”
appears next to the IQT key.

■

Use Central Dial to select CALIBRATE.

■

Close cabinet door.

■

To start calibration press PROG/START and TEMPERATURE key
simultaneously.

■

Calibration starts automatically and lasts 45-60 minutes.

■

After calibration, unit must be switched off for 5 seconds before using again.

■

During calibration, several values will be shown on the display.

■

Once calibration begins, do not interrupt the process.

■

Do not open door until calibration is complete.

After calibration, very hot steam–around 392°F (200°C)–remains in the cabinet.
When opening the cabinet door hot steam will escape, and burns could result.

62

Model

Terminal Positions

LCG-6

5 and 6

LCG-10

5 and 6

LCG-1020

5 and 6

LCG-20

6 and 7

LCG-40

5 and 6