Para 5-34, Para 5-35, Paragraph 5-35 – HP TM 11-6625-2779-14&P User Manual

Page 47: Refer to paragraph 5-34, To paragraph 5-34, See paragraph 5-35

Advertising
background image

Model 3555B

Section V

Table 5-6. Front Panel Trouble Analysis (Cont'd)

3555B SHOULD INDICATE

SET ACTUALLY

CORRECTIVE

INPUT CONDITIONS

FUNCTION

(RANGE + METER)*

INDICATES

ACTION

LEVEL -30dBm

Change RANGE to -30dBm

-30dBm

±

0.2dBm

See Paragraph 5-35

LEVEL 40dBm

Change RANGE to 400dBm

-40dBm

±

0.2dBm

See Paragraph 5-35,

Table 5-9

LEVEL -500dBm

Change RANGE to -50dBm

-50dBm

±

0.2dBm

See Paragraph 5-35,

Table 5-9

LEVEL -60dBm

Change RANGE to -60dBm

-60dBm

±

0.2dBm

See Paragraph 5-35,

Table 5-9

LEVEL -70dBm

Change RANGE to -70dBm

-70dBm

±

0.2dBm

See Paragraph 5-35,

Table 5-9

LEVEL -80dBm

Change RANGE to -80dBm

-80dBm

±

0.2dBm

See Paragraph 5-35,

Table 5-9

LEVEL 0dBm

RANGE to 0dBm

0dBm

INPUT: TMS, TERM

Measure 270mV ac

±

0.2dBm at

AC MON jacks

See Paragraph 5-38

NG CHECK
75 UNBAL, Connect

RANGE: 0dBm

Adjust oscillator

UNBAL signal

FUNCTION: VF/Nm

level for 0dBm

between tip and ring

600 BAL

on 3555B meter

Change input

Depress NG button

-40dBm

Refer to Table 5-8

connection. Connect

Change RANGE to 40dBm

signal between tip
and ring and sleeve

*Some meter jitter may

(tip and ring shorted

be experienced, but the

together), ground

reading should be within

lead to sleeve

the tolerance indicated.

c.

Schematics-- The schematic diagrams
contain dc voltage levels and signal levels
for a specified input condition. This will
assist in troubleshooting individual circuits.

5-31.

FRONT PANEL TROUBLESHOOTING.

5-32.

Before attempting to troubleshoot the set, first

determine from the front panel controls exactly which
functions are performing properly and which ones are
not. In this way, many troubles can be isolated to a
specific area and sometimes to a component.
5-33.

Table 5-6 is a step by step procedure for

checking out the front panel controls. This table
indicates what the results should be for each check
along with the specified tolerance. A space is provided
to enter your results. If these spaces are completed for
each check, they will be of great assistance in making
further troubleshooting checks. Whenever a
discrepancy exists between your results and those

indicated in column 3, refer to the "corrective action"
column.

NOTE

This table is designed to help locate
catastrophic failures. If your set is
only out of the specified tolerances,
a complete adjustment and
calibration procedure should be
performed as described in Paragraph
5-14.

5-34.

FUNCTION TROUBLESHOOTING.
a.

First determine from the Front Panel
Analysis chart (Table 5-6) exactly which
function is defective. Refer to Table 5-7 for
the probable cause of the malfunction in
the FUNCTION switch assembly.

5-35.

RANGE TROUBLESHOOTING.

5-13

Advertising