Connecting to a hart transmitter – Fluke HART 744 User Manual

Page 11

Advertising
background image

Documenting Process Calibrator HART Mode

Connecting to a HART Transmitter

3

Connecting to a HART Transmitter

To connect to and begin to communicate with a HART
transmitter, refer to Figure 1 and proceed as follows:

Note

If you only want a communication connection to
a transmitter that is powered in a loop, simply
clip the HART alligator clips to the loop power
terminals on the transmitter, press

r

, and you

are done.

1. Connect the calibrator mA jacks to the loop power

terminals of the transmitter.

2. Plug the HART interface cable into the SERIAL

PORT, then connect the alligator clips to the same
terminals as in step 1. There is no right or wrong
polarity.

3. Press

the

r

key.

Note

The square-root symbol on

r

key is there

only for calculator mode. At all other times, there
is no square-root function associated with the

r

key.

4. If the transmitter is not powered by a loop power

supply, press the

Loop Power

softkey to activate 24V

loop power.

Note

The calibrator provides loop power through an
internal series resistance of 250

.

If the calibrator shows a measurement of 0 mA,
check for reversed current leads.

If an external loop power supply is used, there
must be a resistance of between 230

and

270

connected in series with the external loop

supply and the transmitter.

5. The calibrator tries Poll Address 0 (single transmitter

per loop). If no connection is made, press the

Poll

softkey to scan Poll Addresses 1 through 15
(multidrop).

6. Once the calibrator establishes communication with

the transmitter, the Active Device screen appears. In
the case of a multidrop configuration, you must
choose a transmitter from a list, and press

e

.

Advertising