Understanding hart calibration, 744 hart mode menus for adjustment – Fluke HART 744 User Manual

Page 21

Advertising
background image

Documenting Process Calibrator HART Mode

Understanding HART Calibration

13

Understanding HART Calibration

An analog transmitter has one stage of electrical
conversion from a measured physical parameter to a
4-20 mA current loop output. A HART transmitter has the
three stages shown in Figure 3.

Depending on how the transmitter is used in your
application, you may need to test and adjust the Input
stage, the Output stage, or both. For example, if your
application requires the Primary Variable (PV) to be
correct when read by a host computer, you must calibrate
the Input stage.

If your application requires that the 4-20 mA current
output value accurately reflect what the Input stage is
measuring, you must calibrate both the Input and Output
stages.

Transmitters in multidrop systems, in which more than
one is wired in parallel, do not use their Output stages at
all. Their analog outputs are all held at an idling level of 4
mA no matter what the Input stage is measuring.

744 HART Mode Menus for Adjustment

In 744 HART mode, adjusting the Input stage is called

Sensor Trim

. Adjusting the Output stage is called

Output

Trim

. Both adjustments are made from the

Service

menu.

For pressure transmitters, there is another adjustment,
called

Pressure Zero Trim

. This adjustment is the same as

trimming the lower sensor point at zero. All three
operations are run from the 744 HART mode

Service

menu.

Advertising