Documenting Process Calibrator HART Mode

Calibrating a Supported HART Transmitter

15

Calibrating a Supported HART
Transmitter

An “as found” and “as left” transmitter calibration has an
easier, more automated process for HART transmitters
than for analog transmitters. The calibration procedure is
the same as described in the

744 Users Manual, except

for how you set up the calibration template, and how you
adjust the transmitter.

Note

If you are starting a calibration procedure from a
loaded Task, connect to the HART transmitter
first and establish communication before you
press

Task

in analog mode.

The following procedure assumes that you are familiar with
using a 740 Series Documenting Process Calibrator to
calibrate analog transmitters, and are not running a loaded
Task.

1. Make the appropriate measuring, sourcing, and HART

interface connections between the calibrator and the
transmitter, and press

r

to establish

communication.

2. Press

r

to switch to analog mode.

pe14s.bmp

3. You are presented with a set of choices for analog

mode. Use the

u

and

d

keys to select one of the

measure/source choices. This is where you select
measure mA (analog output), or measure PV, which
does not involve the Output stage of the transmitter.
See Figure 3.

4. Press

e

.

5. Press the

As Found

softkey.

6. Use the

u

and

d

keys followed by

e

to select

an instrument calibration procedure. You’ll see that
the calibration template is preloaded with appropriate
data. You can make changes to any item if you wish.