Loop test – Fluke HART 744 User Manual

Page 24

Advertising
background image

744
HART Mode Users Guide

16

7. Proceed with calibration as usual (described in the

744 Users Manual).

8. Press

the

Adjust

softkey. This returns you to HART

mode, and presents you with the Service menu for the
transmitter.

9. Do an Output Trim, a Sensor Trim, or both. These are

Service Operations, and are described later in this
manual. When you have finished with the trim
procedures, press the

Done

softkey.

10. Press the

As Left

softkey and repeat the calibration

procedure to verify that the transmitter passes.

Loop Test

The Loop Test feature sends a command to the transmitter
to set its output stage to a specified value. Use this to
check the calibration of the output stage, or to check for
proper reading on an external loop reading device.

To perform a Loop Test:

1. From the Active Device screen Press the

Service

softkey, followed by

Loop Test

.

pe15s.bmp

2. Press a softkey to command the transmitter to set its

analog output to the corresponding value, or manually
enter a value with the numeric keys.

In the middle of the screen the transmitter acknowledges
the selected PVAO (digital representation of the Output
stage). The calibrator shows the measured value at the top
of the screen, so you can use Loop Test as a quick way to
check the calibration of the transmitter’s Output stage.