Field verification of accuracy, Infrared (irt) and probe channels (rtd), Cold temperature verification check – Fluke FoodPro Plus Food Safety Thermometer User Manual

Page 14

Advertising
background image

FoodPro Plus
Users Manual

10

contents of delicases, chilled salads and desserts, holding ovens,
rotisseries, etc.

Note

If any questionable temperatures are indicated or if temperature
readings are within ±1 °C (±2 °F) of the HACCP “Danger Zone”
boundary, use the probe to check internal temperatures.

Field Verification of Accuracy

Infrared (IRT) and Probe channels (RTD)

Follow the steps below to verify the accuracy of your food safety
thermometer. The 0

°

C (32

°

F) “stirred ice bath” is the recommended

verification reference point. Since it is more difficult to measure the surface
of hot water, the hot verification should be used only as a general accuracy
check of the IRT channel.

Cold Temperature Verification Check

1.

Fill a large styrofoam cup halfway to the surface with ice cubes.
Add cold water to just below the rim of the cup.

2.

Immerse the tip of a known calibrated probe thermometer
(reference probe) into the water and stir the mixture with the
probe for one minute, or until the probe temperature stabilizes.

3.

Continue stirring the water with a straw or swizzle stick while
taking simultaneous temperature measurements with the
reference probe and the IR thermometer. The unit should be
held within 3 inches of the surface of the water (Figure 8). To
insure measurement accuracy the probe tip must be immersed
to a minimum depth of 12 mm (1/2 ").