Noncontact Food Safety Thermometer

Field

Verification of Accuracy

13

ebq13.eps

Figure 10. Noncontact Thermometer Working Range

Operating and Ambient Temperature Range

The food safety thermometer is designed to operate in environments
between 0 °C – 50 °C (32 °F – 122 °F). Avoid subjecting the unit to extreme
or abrupt changes in ambient temperatures. Allow the unit to stabilize for a
minimum of 30 minutes if exposed to rapid temperature changes. Failure to
precondition the instrument to the ambient temperature may result in
measurement errors (Figure 11).

Hot to Cold and Cold to Hot

30 Minutes

ebq14.eps

Figure 11. Ambient Temperature Ranges

Response Time

The response time of the instrument from initial start up is less than 1
second. The temperature display is updated at approximately 2 times per
second during sustained operation.