Troubleshooting, Other operational considerations – Fluke FoodPro Plus Food Safety Thermometer User Manual

Page 21

Advertising
background image

Noncontact Food Safety Thermometer

Setting/Changing ºC and ºF Temperature Scales

17

Note

Each time the battery is installed or replaced, the temperature
scale will default to the Celsius scale. To select the Fahrenheit
scale see ‘Setting/Changing

°

C and

°

F Temperature Scales’.

Troubleshooting

Code:

--- (on display)

Problem:

Target temperature is over or under range

Action:

Select target within range specifications

Code:

Battery Symbol

Problem:

Possible low battery

Action:

Check and/or replace battery

Code:

Blank display

Problem:

Dead battery or unit failure

Action:

Check and/or replace battery. If the battery is OK, then
contact a Fluke service center for repair.

Other Operational Considerations

All models are protected from the following:

EMI (Electro Magnetic Interference) from induction heaters and

microwave ovens

Electrostatic discharge

Should the unit become damaged, check the accuracy of the unit by

performing the verification process recommended in this manual. If the
unit is out of calibration, do not rely on it for critical temperature
measurements. Contact a Fluke service center for repair,

Heat from stovetops, pans, or other hot surfaces (don’t set it on the

stove)

Advertising