Fluke 1522 User Manual

Page 50

Advertising
background image

Description

Command

Example

Response
Format

Response
Example

Range

Set autolog mode on/off .

Note: Exit Autolog menu

to use this command. After
alog =ON is sent, the buttons
are disabled until the command
alog=OFF is sent.

alog=[ON/OFF]

alog=ON

ON or OFF

Clear autolog memory

aclr

aclr

Initiate download of autologged
data.

Note: Read Section

11.2 on downloading
autologged data before at-
tempting to use this
command.

daut

daut

Read autolog interval

aint

aint

int: 99 {secs or min}

int: 60 {secs or min}

Set autolog interval

Note: See Section 7.4.2

for specifics on log interval.

aint=n

aint=30

1, 5, 10, 10,
15, 30, 60,
120, 300, 600,
900, 1800,
3600

Read delta(x) measurement

dx

dx

dx: 999.999
{C, O, F, R, or K}

dx: -0.118 C

Set delta(x) reference to pres-
ent measurement

sd[x]

sd

Read unit

u[nit]

u

u: {C, O, F, R, or K}

u: C

Set unit

u[nit]=C/O/F/K/R

u=C

C, O, F, K, R

Read serial sample rate

sa[mple]

sa

sa: 99999

sa: 10

Set serial sample rate

sa[mple]=

n

sa=60

0 to 10,000

Read resolution digits

re[solution]

res

res: 9

res: 3

Set resolution digits

re[solution]=

n

res=0

1 to 3

Read time

ti[me]

time

ti: hh:mm:ss

ti: 16:23:45

Set time

Note: You must set

hours, minutes, and seconds
for the time to be set
correctly.

ti[me]=hh:mm:ss

ti=08:15:00

00:00:00 to
23:59:59

Read identification information

*i[dn?]

*idn?

<manufacturer>,
<model>,
<serial number>,
<firmware version>

Hart,
1522
95001,
1.24

Read model and firmware
version

*v[ersion]

*ver

ver.<model>,
<firmware version>

ver.1522,
1.24

Legend:

[] Optional Command data

{} Returns either information

49

8 Communications Interface

Communication Command List