2 calibration commands 50, Calibration commands, 2 calibration commands – Fluke 1522 User Manual

Page 51

Advertising
background image

Description

Command

Example

Response
Format

Response
Example

Range

n Numeric data supplied by user

9 Numeric data returned to user

x Character data returned to user

8.2.2

Calibration Commands

The following group of commands directly or indirectly affect the accuracy of
the instrument. These commands should only be used by qualified, authorized
personnel. Calibration parameters can only be changed if protection is dis-
abled by entering the proper pass-code using the command *pas= ddddd.
Entering an invalid passcode or cycling the power re-establishes protection of
the calibration parameters. The parameters can be read without disabling
passcode protection.

Description

Command

Example

Response
Format

Response
Example

Range

Unprotect calibration parame-
ters with the pass-code

*pas[scode]=nnnnn

*pas=12345

Read passcode protection
state

*pas

*pas

ps: 9

Ps: 0

0 or 1

Read date

da[te]

date

da: yyyy-mm-dd

da: 1999-03-28

Set date

da[te]=yyyy-mm-dd

da=1999-03-28

1999-01-01 to
2100-12-31

Read probe serial number

se[rial]

ser

ser: xxxxxxxxxx

ser: 100231A

Set probe serial number

se[rial]=xxxxxxxxxx

ser=001

0 to 9, A to Z,
., -, /

Read probe type

p[robe]

pr

pr: {ITS-90, IEC751,
CVD, YSI400, THERM}

pr: ITS-90

Set probe type

p[robe]=IT[S-90]/
IE[C751]/ C[VD]/
Y[SI400]/ T[HR]

pr

ITS-90,
IEC751,
CVD, YSI400,
THERM

Read ITS-90 R(.01) or CVD R0 r[0]

r0

r0: 999.999

r0: 100.035

Set ITS-90 R(.01) or CVD R0

r[0]=n

r0=100.106

1 to 900

Read ITS-90 coefficient

a/b/c/d/a4/b4

a

{a/b/c/d/a4/b4}:
-9.999999E-99

a: -2.8644101E-05

Set ITS-90 coefficient

a/b/c/d/a4/b4=n

a=-2.8644E-5

1.0 to 9.9e-18

Read CVD coefficient

al[pha]/de[lta]/be[ta]

al

{al/de/be}: 9.99999999 al: 0.00385762

Set CVD coefficient

al[pha]/de[lta]/be[ta]=n al=0.0038512

Coefficient
Dependent

Read THERM thermistor
coefficient

b0/b1/b2/b3

b3

{b0/b1/b2/b3}:
9.999999E-99

b3: -9.944070E-12

Set THERM thermistor
coefficient

b0/b1/b2/b3=n

b3=-9.944E-12

E+/-18

1522 Handheld Thermometer Readout
User’s Guide

50