Approvals, Specifications – Fluke 1552A EX User Manual

Page 12

Advertising
background image

1551A Ex/1552A Ex

Calibration Manual

4

Approvals

The Product meets the above requirements using:
EN 60079-0:2006 ELECTRICAL APPARATUS FOR
EXPLOSIVE GAS ATMOSPHERES -- PART
0: GENERAL REQUIREMENTS (IEC
60079-0:2004 (MOD))

EN 60079-11:2007
EXPLOSIVE ATMOSPHERES -- PART 11:
EQUIPMENT PROTECTION BY INTRINSIC
SAFETY "I" (IEC 60079-11:2006 (EQV) +
CORRIGENDUM DEC. 2006 (EQV))

Manufactured by Fluke Corporation:
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.

P.O. Box 1186
5602 BD Eindhoven
The Netherlands

Specifications

(Ambient: 23

°

C

±

5

°

C)

Measurement Range

1551A

Ex

.........................................................................

50

°

C to 160

°

C (

58

°

F to 320

°

F)

1552A

Ex

.........................................................................

80

°

C to 300

°

C (

112

°

F to 572

°

F)

Accuracy (1 year) .....................................................................

±

0.05

°

C (0.09

°

F)

Resolution ................................................................................. Selectable (0.1, 0.01, 0.001) factory default is 0.01
Sample Rate ............................................................................. User selectable 0.5/sec, 1/sec or 2/sec factory default is

1/sec

Operating Temperature Range of Readout: ..........................

10

°

C to 50

°

C (14

°

F to 122

°

F)

Probe Response Time .............................................................. Approximately 20 seconds
EMC Compliance....................................................................... EN61326:2006 Annex C

CISPR 11, Edition 5.0-2009

Class

“B”

Humidity Range......................................................................... 0 to 95 % RH Non-condensing

Storage Temperature Range....................................................

20

°

C to 60

°

C (

4

°

F to 140

°

F)

Enclosure Rating ...................................................................... IP50

Readout Temperature Coefficient ........................................... Add

±

10 ppm/

°

C of full scale temperature from

10

°

C to

18

°

C and 28

°

C to 50

°

C

Probe Temperature Coefficient .............................................. 0.00385

Ω

/

Ω

/

°

C nominal

Advertising