Record as-found calibration parameters, Evaluate as-found measurement errors – Fluke 1552A EX User Manual

Page 15

Advertising
background image

Stik Thermometer

Calibration

7

Record as-found Calibration Parameters

Before any adjustments are made, the present values of the Product calibration

parameters should be recorded. Connect the serial cable between the computer and the

Product. The commands used to read the values of the calibration parameters are given in

Table 3. Commands must be terminated with a carriage return character. Table 3 may

also be used to record the values of the calibration parameters.

Table 3. As-found Calibration Parameters

Parameter

Command

Value (fill in)

TEMP1 CAL:USER:TEMP1?

ADJ1 CAL:USER:ADJ1?

TEMP2 CAL:USER:TEMP2?

ADJ2 CAL:USER:ADJ2?

TEMP3 CAL:USER:TEMP3?

ADJ3 CAL:USER:ADJ3?

DATE CAL:DEV:DATE?

Evaluate as-found Measurement Errors

The Product should be tested to evaluate its accuracy before any adjustments are made. It

is recommended that it be tested near the five temperatures listed in Table 4, depending

on the model.

Table 4. Test Temperatures

Point

1551A Ex

1552A Ex

1 -50

°

C -80

°

C

2 -25

°

C -25

°

C

3 0

°

C 0

°

C

4 100

°

C 100

°

C

5 157

°

C 300

°

C


At each of the test temperatures, readings are recorded of the temperature of the reference

thermometer and the indication of the Product. Before recording the readings, make sure

the temperatures of the bath, reference thermometer, and 1551A Ex/1552A Ex

Thermometer are settled and stable.