1551A Ex/1552A Ex

Calibration Manual

6

Recommended Equipment

The equipment listed in Table 2 will satisfy the calibration requirements:

Table 2. Recommended Equipment

Equipment Recommended

Model

Temperature Baths

Fluke 6331 40

°

C to 300

°

C Hot Bath

Fluke 7381 -80

°

C to 110

°

C Cold Bath

Fluke 7196 Equilibration Block

Reference Thermometer

Fluke 5699-S Extended Range Metal-Sheath SPRT

Reference Thermometer Readout

1594A Super-Thermometer readout

Computer

Computer with an RS-232 com port
RS-232 serial cable
Terminal software to communicate with the 1551A

Ex/1552A Ex Thermometer

Initial Setup

Check the batteries in the Product and replace them if necessary. The battery level

indication is found in the Setup menu.
Several settings in the Setup menu relate to measurement and should be set as follows:

Sample Rate:

1 per second

Damping:

OFF

Display Resolution:

0.001

Auto Power Off:

OFF

The unit of measurement in the 1551A Ex/1552A Ex Thermometer may be set to °C or

°F, but must match the unit of the reference thermometer.
The baud rate in the computer software must match the baud rate of the 1551A Ex/1552A

Ex Thermometer, which is usually 9600 but may be set to 2400.