Set the calibration date, Evaluate as-left measurement errors – Fluke 1552A EX User Manual

Page 18

Advertising
background image

1551A Ex/1552A Ex

Calibration Manual

10

The command will set the value of the CAL:USER:TEMP3 parameter to the given

temperature and automatically adjust the CAL:USER:ADJ3 parameter to cause the

Product to measure the given temperature accurately. The adjustment made here will not

affect the measurements at the other Adjustment temperatures.
Allow the readings of the Product to settle and then verify that the readings match the

temperature of the reference thermometer well within the specified accuracy of the

1551A Ex/1552A Ex.

Set the Calibration Date

The date of calibration should be stored in the Product using the following command:

CAL:DEV:DATE <year>,<month>,<day>

Parameters <year>, <month>, and <day> are numbers representing the year, month, and

day of the month respectively.
After all the adjustments have been made and the calibration date has been set, protection

of the calibration parameters should be restored using the following command:

SYST:PASS:CDIS

(Protection will always be restored when the Product is switched on.)

Evaluate as-left Measurement Errors

After all adjustments have been made, the Product should be subjected to a final test to

ensure that it is measuring accurately over its entire range. The measurements are

collected and errors are evaluated in the same way that was explained in “Evaluating As-

found Measurement Errors.” Table 9 may be used to record the as-left data.

Table 9. As-left Measurement Errors

Point

Reference (fill in)

1551A Ex/1552A Ex (fill in)

Error (fill in)

1

2

3

4

5

Record as-left Calibration Parameters

As the last step in the calibration procedure, the final values of the Product calibration parameters
should be recorded. The commands required to read the values of the calibration parameters are
given in Table 10. Each command must be terminated with a carriage return character. The table
may also be used to record the values of the calibration parameters.

Table 10. As-left Calibration Parameters

Parameter

Command

Value (fill in)

TEMP1 CAL:USER:TEMP1?

ADJ1 CAL:USER:ADJ1?

TEMP2 CAL:USER:TEMP2?

ADJ2 CAL:USER:ADJ2?

TEMP3 CAL:USER:TEMP3?

ADJ3 CAL:USER:ADJ3?

DATE CAL:DEV:DATE?