Concentrator maintenance – Invacare IRC5LXO2 User Manual

Page 18

Advertising
background image

18

PART

ONE

Cabinet Filter

(located on both sides

of concentrator)

NOTE: There are two (2) cabinet filters. One (1) is located on each side of

the cabinet (FIGURE 1).
1. Remove each filter and CLEAN AT LEAST ONCE A WEEK

depending on environmental conditions (FIGURE 2).

CONCENTRATOR MAINTENANCE

MAINTENANCE

FIGURE 1 - CABINET FILTER LOCATION

NOTE: The Invacare Concentrators are specifically designed to minimize

routine preventive maintenance at intervals of once per year. In places with high

dust or soot levels, maintenance may need to be performed more often. The

following must be performed at a minimum of one (1) year in service to assure

years of additional reliability. Only authorized or factory-trained personnel should

perform preventive maintenance on the concentrator. Power should be discon-

nected before beginning preventive maintenance on the Invacare concentrator.

CLEANING THE CABINET FILTERS

NOTE: Environmental conditions that may require more frequent cleaning

of the filters include but are not limited to: high dust, smoking, air pollutants, etc.
2. Clean the cabinet filters with a vacuum cleaner or wash in

warm soapy water and rinse thoroughly (FIGURE 2).

3. Dry the filters thoroughly before reinstallation.

CAUTION

DO NOT operate the concentrator without the filters installed.

FIGURE 2 - CLEANING THE CABINET FILTERS