Troubleshooting – Invacare IRC5LXO2 User Manual

Page 20

Advertising
background image

20

PART

ONE

TROUBLESHOOTING

1. Internal repairs required. 1. Call Service Provider.

GREEN LIGHT with

YELLOW Light

Flashing.

Alarm:

RAPID
Beep....Beep...

Beep...Beep

LOW FLOW ALARM

1a. Kinked or blocked

tubing, cannula or

humidifier.

1b. Flowmeter set at

0.5 L/min. or less.

1a. Inspect for kinks or

blockages. Correct,

clean or replace item.

Once corrected, turn

power OFF for 60

seconds and then turn

power back ON.

1b. Reset flowmeter to

prescribed flowrate. If

prescribed flowrate is

below 0.5 L/min., use

the Pediatric

Flowmeter IRCPF16.

SYMPTOM:

YELLOW or RED

LIGHT Illuminated.

Alarm:

CONTINUOUS
On red light only.

PROBABLE CAUSE:

1. Low oxygen purity.*
2. Kinked or blocked

tubing, cannula or

humidifier.*

3. Flowmeter set at

1/2 L/min. or less.*

4. Unit overheating

due to blocked air

intake.

5. Insufficient power at

outlet.

6. Internal repairs required.

SOLUTION:

1. Clean or Replace filters.
2. Inspect for kinks or

blockages. Correct,

clean or replace item.

Once corrected, turn

power OFF for 60

seconds and then turn

power back ON.

3. Check flowmeter is set

to 1.0 L/min. or more.

Refer to Recommended

End-User Guidelines in

the Operating

Instructions of this

manual.

4.a.Remove and clean

cabinet filters.

4b. Move concentrator at

least 3- inches from

walls, draperies, and

furniture.

5. Do not use extension

cords. Move to another

electrical outlet or

circuit.

6. Call Service Provider.

* Only applies to

SensO

2

models.