Concentrator troubleshooting guide – Invacare IRC5LXO2 User Manual

Page 19

Advertising
background image

19

PART

ONE

CABINET

WARNING

UNPLUG the concentrator when cleaning. To avoid

electrical shock, DO NOT remove cabinet.

1. Clean the cabinet with a mild household cleaner and non-abrasive

cloth or sponge.

HUMIDIFIER

NOTE: To clean the oxygen humidifier, follow the instructions provided by

the manufacturer. If none are provided, follow these steps:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten (10) parts

water and one (1) part vinegar.

3. Rinse thoroughly with hot water and refill with distilled water to

the level shown on the humidifier.

WARNING

DO NOT overfill Humidifier Bottle.

SOLUTION:

1. Insert plug into outlet.
2. Inspect house circuit

breakers or fuses. If problem

recurs, use a different outlet.

3. Call Service Provider.

1a. Remove and clean

cabinet filters.

1b. Move oxygen concentrator

at least three (3) inches

away from walls, draperies

or furniture.

2. Do not use extension

cords. Move to another

electrical outlet or circuit.

3. Call Service Provider.

CONCENTRATOR

TROUBLESHOOTING GUIDE

PROBABLE CAUSE:

Main Power Loss:

1. Power cord NOT

plugged In.

2. No power at outlet.
3. Tripped circuit breaker.

System Failure:

1. Unit overheating due

to blocked air intake.

2. Insufficient power at

outlet.

3. Internal repairs

required

SYMPTOM:

Alarm:

SHORT BEEPS,

LONG PAUSE
Concentrator

NOT operating,

power switch ON.
Beep......Beep......
Alarm:

CONTINUOUS

Concentrator

NOT operating,

Power Switch ON.

Beeeeeeeep....

MAINTENANCE/TROUBLESHOOTING