Turning the compressor on (figure 3) – Invacare IRC5LXO2 User Manual

Page 64

Advertising
background image

16

PART

TWO

COMPRESSOR OPERATION

PROCEDURE 3

TURNING THE COMPRESSOR ON

(FIGURE 3)

NOTE: Refer to the GENERAL WARNINGS and the HANDLING WARNINGS

in the SAFETY SUMMARY of this manual.
1. Make sure the concentrator is on. Refer to the concentrator

owner’s manual, Part One.

CAUTION

The concentrator flow rate to the patient MUST be at 3 L/

min. or less. Otherwise the oxygen level to the compressor

will be <90% and the compressor will not fill the cylinder.

2. Set the concentrator flow rate to the patient to 3 L/min. or less.

Refer to the concentrator owner’s manual, Part One.

3. Make sure the cylinder is connected to the compressor. Refer to

CONNECTING THE CYLINDER TO THE COMPRESSOR in this

section of the manual.

4. Push the power switch on the control panel to the ON position.
5. Examine the control panel. The following sequence of events

should occur:
A. 0-3 MINUTES -

The WAIT (YELLOW) light will be lit while the compressor

warms up.

B. AFTER 3 MINUTES -

The FILLING (GREEN) light will be on while the cylinder is

filling.
The FULL (GREEN) light will be on when the cylinder is finished

filling. Proceed to STEP 6.

NOTE: The O2 BELOW NORMAL (YELLOW) light may come on if the

oxygen level from the concentrator has not yet reached or drops below 90%.

The compressor will begin or resume filling when the oxygen level reaches 90%.

If light DOES NOT go off within 10 minutes, refer to the TROUBLESHOOT-

ING section of the Homefill II Owner’s Manual (Part No. 1100873).