4. Maintenance


The osmosis treatment unit runs completely
automatically and maintenance is minimal. The
only essential operation is periodic replacement
of the filter cartridges.

IMPORTANT
Replace the pre-filter every 3000 hours of
operation and AT LEAST 1 time a year.






















Pre-filter Membrane

Replacing the pre-filter:

Filter cartridge replacement procedure

-

Shut off power;

-

Close the side cocks (RL);

-

Vent the circuit from the top cap (S);

-

Unscrew the filter housing (use the special wrench supplied in the
accessories bag). Place a container to drain the water in the filter
into;

-

Fit the new cartridge and check it is correctly seated;

- Close

vent

S and open the side cocks RL;

-

Check for leaks and water in the bottom of the appliance;

-

Dry it off thoroughly.

Replacing the membranes:
Cartridge replacement procedure.

-

Shut off power;

- Disconnect all hoses from the membrane

housings, unscrew the base, and place a
container to catch the water contained in the
cartridges;

-

Refit the membrane, check the seating of the
o-ring and screw the housing fully in;
reconnect the hoses and make sure they are
correctly seated;

-

Do not apply excessive force;

-

Fit the new cartridge and check it is correctly
seated;

-

Start the unit up and check for leaks.