B) osmosis-treated water not produced – Smeg WO-01 User Manual

Page 17

Advertising
background image

-

43

-

B) Osmosis-treated water not produced

Possible

causes

Subcause Solution

Display

alarm

message

Check power supply and power switch ON
(see “Positioning and installation”)

No electrical power

Power plug not inserted in electrical panel
(insert)
Check connections and whether board is
sending a signal

No power to pump

Pump condenser faulty (replace)

Intake pressure switch faulty (jammed on full)

Replace, or adjust the pressure with the
adjuster screw (fig.5)


Vat level switch (low) faulty (jammed on full)

Max level switch faulty, does not send
production enable signal (replace/clean)
(see wiring diagram)

none

No water supply or supply pressure/flow
inadequate

- Supply cock closed OPEN;

- Supply filter blocked CLEAN

- Supply pressure should be > 1.5 bar
- Pre-filter up/downstream cocks closed
OPEN (see maintenance section)

- Filter blocked – REPLACE (fig.5)

- Filler solenoid valve faulty or no power –
REPLACE IF FAULTY (fig.5)

- Intake pressure switch faulty (jammed on
empty) – REPLACE (fig.5)

Low pressure

Osmosis pump
not running

Water in vat


Vat level switch (high) faulty, does not
switch to full: continues producing with
overflow. REPLACE (see wiring diagram)

Flooding

alarm

Water supply too cold (<10°)

Increase supply temperature

Pre-filter blocked

Replace pre-filter (see maintenance
section)

Membranes blocked

Replace membranes (see maintenance
section)

Impeller damaged/broken

Check pressure gauge reading
if > 8, replace pump or impeller as
required

Bypass valves need adjustment

See special adjustments section

Discharge pipe crimped

Route pipe properly

Osmosis pump
running but
production low

Pressure drop in vat supply hoses

Shut off power, dry thoroughly and repair
leak