List of faults, A) no osmosis-treated water delivery, Faults – Smeg WO-01 User Manual

Page 16

Advertising
background image

-

42

-

6- Faults

Front door leds

The door has two leds to indicate:

- Green: power on, normal operation;
- Red: fault, production of osmosis-treated water stopped

Important:

THE DOOR GIVES NO INDICATION OF DELIVERY
CIRCUIT FAULTS
See below for details.


Green



Red

List of faults

A) No osmosis-treated water delivery

Possible causes

Subcause

Solution

Display alarm

message

No water in vat

See point "B"

see point "B"

Check power supply and power switch ON
(see “Positioning and installation”)

No electrical power

Power plug not inserted in electrical panel
(insert)
Check connections and whether board
is sending a signal

No power to pump

Pump condenser faulty (replace)

Vat pressure switch faulty (jammed on
empty)

Empty vat (manually with jumper on delivery
pump or cock –fig.4) and replace pressure
switch (fig.1)

Delivery pressure switch faulty (jammed
on full)

Replace, or adjust the pressure with the
adjuster screw (fig.2)

Non-return valve jammed (does not
open)

Check valve, if jammed replace/clean (fig.3)

Delivery pump not
running

Vat not full enough for use (min level)

Wait for a few minutes while it fills

No water arriving at pump: vat suction
pipe blocked/crimped/cock closed

Remove obstacle/open cock (fig.4)

Air in delivery pipe

Shake pipe and see if air rises into the vat

Impeller damaged/broken

Replace

The user requires minimum intake
pressure of >1.8 BAR

System limit

Delivery pump runs
but does not deliver
or only insufficiently

The user requires flow > 20 l/min

Reduce user load

none