Sterilite Point-of-Use Drinking Water System User Manual

Page 10

Advertising
background image

10

It is EXTREMELY IMPORTANT that the unit

be unplugged before servicing. The Sterilight

Point-of-use unit contains a high intensity

ultraviolet lamp that can damage tissues in

the eye. The UV wavelengths cannot travel

through the plastic housing. However, the

lamp will be exposed during removal and

replacement. In addition, it is possible to

encounter hazardous electrical energy when

removing the protective bottom cover.

DO NOT PROCEED WITHOUT UNPLUGGING

THE UNIT!

1. Disconnect unit from power source.

2. Remove the unit from its wall mounting,

remove the filter, drain the water from the

unit (Figure 21), and lay it down horizontally.

3. Using a Phillips-head screwdriver, remove

the retaining screws on the cover located

at the bottom of the unit (Figure 22) and

carefully slide the cover off of the unit .

4. Disconnect the lamp wires.

NOTE: Depending on model, your unit may

have an internal or external power supply.

Units with an internal power supply appear

as in Figure 23. Units with an external power

supply appear as in Figure 24.

Figure 25 (External Power Supply)

Loosening lamp retention screw

Figure 22

Removing cover retaining screws

Figure 23 (Internal Power Supply)

Removal of Single Socket

lamp Connection

Figure 24 (External Power Supply)

Figure 21—Draining the Unit