26

SSCI - Suburban Surgical Company, Inc.

Chapter 4 - Repairs and Replacements

The following instructions cover both the seal strip retainers and the
seal strips. All seal strip retainers are the same except for length and
number of mounting holes. The same is true for all seal strips.

Parts Identification
Table 3 on Page 27 and Table 4 on Page 28 give the part numbers
for the various seal strip retainers and seal strips.
Note: The dimensions refer to the door/cage size, and not to the
actual lengths of the retainers and seals.

Tools Required
„

3/8 in. box-end or open-end wrench

„

Phillips screwdriver

Procedure
Note:
You do not have to remove the door from the cage to
replace a seal strip retainer or a seal strip.

1.

Open the oxygen therapy door.

2.

With a 3/8 in. wrench and a Phillips screwdriver, remove
the mounting nuts and screws (Figures 29 and 30) that hold
the seal strip retainer and seal strip to the door.
Note: The number of mounting screws/nuts varies with the
dimensions of the door.

Seal Strip Retainer
and Seal Strip

Refer to Tables 3 and 4 for

Part Numbers

3.

Lift off the seal strip retainer.

4.

If you are replacing the seal strip, remove it at this time.

Figure 29. Outside View - Seal Strip and
Retainer Mounting Screws

Seal Strip

Mounting Screw

Figure 30. Inside View - Seal Strip
Retainer and Seal Strip

Mounting

Nut

Seal Strip

Seal Strip Retainer