Make sure that you:

❏ Touch only the tip of

the drop of blood to the
test spot.

❏ Do not smear blood on the

test spot or apply a second
drop after the test begins.

❏ Apply enough blood to form

a round, shiny drop that
covers the test spot com-
pletely and stays wet during
the entire test.

❏ Do not move the test strip

as you are applying blood. If
the test strip moves, push it
back to its original position.

❏ Do not remove the test

strip from the meter to
apply blood.

The meter will beep when the
blood has been applied to the
test spot.

Step 3: Accurate Results in
45 Seconds.

The meter then counts down
from 45 to 0 seconds,
followed by
a series of
beeps when
your result is
displayed.

106

(example)

Be sure to wait for the beeps
before you note your result.
The meter displays results
which are between 0 and
600 mg/dL (0–33.3 mmol/L).
Results above 600 mg/dL
(33.3 mmol/L) are displayed as

high

To select an event number
at this time, see page 65.