Pride Mobility QUANTUM 1121 User Manual

Page 39

Advertising
background image

Quantum 1121

www.quantumrehab.com

39

V I I . O P E R A T I O N

Actuator Lighting Module (Not Shown)

The actuator lighting module is also located on the power base. The actuator lighting module provides a control and power
interface between the power module, the lights, and/or the power seat actuator.

Sleep Mode

Your Remote Plus controller has a sleep mode feature. Sleep mode is a built-in circuit that automatically shuts off the main
power if the joystick is not moved in any direction for approximately five minutes. The battery condition meter lights on the
keypad indicate sleep mode by blinking once every five seconds. To restore power and continue, push the on/off key
twice.

Thermal Rollback

The Remote Plus is equipped with a thermal rollback circuit. This circuit monitors the temperature of the motors, power
module, and remote. In the event that any of them become excessively hot (above 122° F), motor voltage is reduced. For
every degree above 122° F, the voltage is reduced by 5 volts. This reduces your power chair’s speed and allows the
electrical components to cool down. When the temperature returns to a safe level, your power chair resumes its normal
speed.

Remote Plus Trouble Codes

In addition to indicating the current state of battery charge, the battery condition meter can also indicate possible problems
with your power chair. The battery condition meter has ten lights. The lights provide information by the number of lights that
are flashing. If any of the meter lights are flashing rapidly, the controller may be indicating a fault. For instance, if the first light
is flashing rapidly, the battery voltage is nearly depleted. The following is a list of the possible errors signified by the rapidly
flashing meter. When you get a trouble code, contact your Quantum Rehab Specialist.

FLASHING LIGHTS

DIAGNOSIS

SOLUTION

10

High Battery Voltage

Check batteries.

9

Solenoid Brake Fault

Check motor/brake wiring.

8

Possible Controller Fault

See Quantum Rehab Specialist.

7

Possible Joystick Fault

See Quantum Rehab Specialist.

6

Inhibit Active

Unplug charger. Check connections.

5

Right Motor Wiring Fault

Check right motor wiring.

4

Right Motor Disconnected

Check right motor wiring.

3

Left Motor Wiring Fault

Check left motor wiring.

2

Left Motor Disconnected

Check left motor wiring.

1

Low Battery Voltage

Check batteries/battery wiring.