Eizo RADIFORCE RX211 User Manual

Page 51

Advertising
background image

Immunity test

EN60601

test level

Compliance level

Electromagnetic environment -

guidance

Conducted RF

EN61000-4-6

3Vrms

150kHz to 80MHz

3Vrms

150kHz to 80MHz

Portable and mobile RF

communications equipment

should be used no closer to any

part of the RadiForce series,

including cables, than the

recommended separation distance

calculated from the equation

applicable to the frequency of the

transmitter.

Recommended Separation

distance

d = 1.2

√P

Radiated RF

EN61000-4-3

3Vrms

80MHz to 2.5GHz

3Vrms

80MHz to 2.5GHz

d = 1.2

√P, 80MHz to 800MHz

d = 2.3

√P, 800MHz to 2.5GHz

Where “P” is the maximum

output power rating of the

transmitter in watts (W) according

to the transmitter manufacturer

and “d” is the recommended

separation distance in meters (m).

Field strengths from fixed RF

transmitters, as determined by an

electromagnetic site survey, should

be less than the compliance level

in each frequency range.

Interference may occur in the

vicinity of equipment marked with

the following symbol.

Recommended separation distance between portable and mobile RF communications equipment and the

RadiForce series
The RadiForce series are intended for use in an electromagnetic environment in which radiated

RF disturbances are controlled. The customer or the user of the RadiForce series can help prevent

electromagnetic interference by maintaining a minimum distance between portable and mobile RF

communications equipment (transmitters) and the RadiForce series as recommended below, according to

the maximum output power of the communications equipment.

Rated maximum output

power of transmitter

W

Separation distance according to frequency of transmitter

m

150kHz to 80MHz

d = 1.2

√P

80MHz to 800MHz

d = 1.2

√P

800MHz to 2.5GHz

d = 2.3

√P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance

“d” in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where

“P” is the maximum output power rating of the transmitter in watts (W) according to the transmitter

manufacturer.

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and

reflection from structures, objects and people.

Cable length
Power Cord :

Accessary

2.0m

Signal Cable (FD-C39) :

Accessary

2.0m

Signal Cable (MD-C87) :

Accessary

1.8m

USB Cable (MD-C93) :

Accessary

1.8m

Advertising