ResMed escape s8 User Manual

Page 36

Advertising
background image

32

R

EPLACING

THE

A

IR

F

ILTER

To replace the air filter:

1. Remove the air filter cover at the back of the

S8 A

UTO

S

ET

S

PIRIT

.

2. Remove and discard the old air filter.
3. Insert a new filter with the blue tinted side facing out from the device.
4. Replace the air filter cover.

Inspect the air filter every month to check if it is blocked by dirt or contains holes. With
normal use of an

S8 A

UTO

S

ET

S

PIRIT

, the air filter needs to be replaced every six months

(or more often if your device is in a dusty environment).

!

WARNING

Do not wash the air filter. The air filter is not washable or reusable.

The air filter cover protects the device in the event of accidental liquid spillage
onto the device. Ensure the air filter and air filter cover are fitted at all times.

S

ERVICING

!

CAUTION

Do not attempt to open the S8 AutoSet Spirit case. There are no user
serviceable parts inside.

Inspection and repair should only be performed by an authorised agent. Under
no circumstances should you attempt to service or repair the flow generator
yourself.

This product (

S8 A

UTO

S

ET

S

PIRIT

) should be inspected by an authorised ResMed

service centre 5 years from the date of manufacture. Prior to this, the device is
intended to provide safe and reliable operation provided that it is operated and
maintained in accordance with the instructions provided by ResMed. Applicable
ResMed warranty details are provided with the device at the time of original supply.
Of course, as with all electrical devices, if any irregularity becomes apparent, you should
exercise caution and have the device inspected by an authorised ResMed service
centre.
If you feel that your flow generator is not performing properly, see “Troubleshooting”
on page 35.

Air filter cover

Advertising