Display error message:

Check tube!!

Key if done

The air tubing is loose or
blocked.

Check that the air tubing is
connected securely to your mask
and the air outlet on the front of
the S8 AutoSet Spirit.
When you have checked the air
tubing, press the Start/Stop key to
restart the device.
If this does not clear the message,
disconnect the power cord and
then reconnect it to restart the
device.

Display error message:

SYSTEM ERROR

Call service!

Component failure.

Return your S8 AutoSet Spirit for
servicing.

Display error message:

High leak in

last session

You have experienced
excessively high leak levels
during the night.

Check that your air tubing is
connected properly and that your
mask does not leak excessively.
Use the mask-fitting feature to
help you to fit your mask
properly.
If this message appears again,
contact your clinician.

Display error message:

HIGH LEAK!!!

Adjust Mask

You have experienced
excessively high leak levels for
more than 20 seconds.

Check that your air tubing is
connected properly and that your
mask does not leak excessively.
Use the mask-fitting feature to
help you to fit your mask
properly.
If this message appears again,
contact your clinician.

Display error message:

Exxxx

(where xxxx

defines an error)

Component failure.

Return your S8 AutoSet Spirit for
servicing.

The following message
is displayed on the
LCD after you try to
update settings using
the Data Card:

Card Error

Remove Card

The Data Card is not inserted
correctly.

Ensure that the Data Card is
inserted with the arrow facing up
and pointing towards the device.
Ensure that the device is switched
on and in standby mode before
you insert the Data Card.

The Data Card is only partly
inserted.

Ensure that the Data Card is
inserted into the slot on the Data
Card module as far as it will go.

Problem

Possible Cause

Solution