Disposal – Omron Professional Blood Pressure Monitor HBP-1300 User Manual
Page 31
29
Mainten
ance
Other problems
Disposal
Description
As there is a risk of environmental pollution, follow your applicable Federal, state and local legal
regulations regarding disposal or recycling of this equipment and batteries.
The main constituents of each part are listed in the table below. As there is a risk of infection, do not
recycle patient attachments such as cuffs, but dispose of them as instructed by your facility's
procedures and applicable regulations.
Error
code
Priority
Cause
Solution
E3
Low
Inflated the cuff to 300 mmHg or more
during inflation in "Auscultation Mode".
When inflating in "Auscultation Mode",
release the button when the pressure
reaches the desired value.
Over-inflation occurs
If this occurs during measurement, repeat
measurement.
If this occurs when not performing
measurement, contact OMRON
HEALTHCARE Customer Service.
E40
Medium The battery is depleted.
Recharge the battery, or replace with a new
battery. (page 14)
E41
Medium Battery failed to charge.
Try charging again. If continues to fail,
replace with a new battery. (page 14)
E42
Medium Battery voltage error
Replace the battery with a new battery. If the
error continues, contact OMRON
HEALTHCARE Customer Service.
Item
Parts
Material
Package
Box
Cardboard
Cushion
Cardboard
Bag
PE
Main unit and
accessories
Enclosure
ABS, PC , SR
Internal parts
General electronic components
Battery pack
Outer tube
PVC
Cell batteries
Nickel-hydrogen battery
Internal parts
General electronic components
HBP-1300.book Page 29 Wednesday, March 7, 2012 9:28 AM