Manufacturer's Declaration

The HBP-1300 is intended for use in the electromagnetic environment specified below.
The customer or the user of the HBP-1300 should assure that it is used in such an environment.
Electromagnetic Emissions:
(IEC60601-1-2)

Electromagnetic Immunity:
(IEC60601-1-2)

Emission Test

Compliance

Electromagnetic Environment

RF emission
CISPR 11

Group 1

The HBP-1300 uses RF energy only for internal functions.
Therefore, this RF emission is extremely weak and there is
little chance of it creating any kind of interference
whatsoever with nearby electronic equipment.

RF emissions
CISPR 11

Class B

The HBP-1300 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.

Harmonic emissions
IEC 61000-3-2

Class A

Voltage fluctuations/flicker
IEC 61000-3-3

Complies

Immunity test

IEC60601-1-2

test level

Compliance level

Electromagnetic

environment - guidance

Electrostatic discharge
(ESD)
IEC 61000-4-2

±6 kV contact
±8 kV air

±6 kV contact
±8 kV air

Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.

Electric fast transient/
burst
IEC 61000-4-4

±2 kV for
power supply lines
±1 kV for
input/output lines

±2 kV for
power supply lines
±1 kV for
input/output lines

Mains power quality should be that of
a typical commercial or hospital
environment.

Surge
IEC 61000-4-5

±1 kV
differential mode
±2 kV
common mode

±1 kV
differential mode
±2 kV
common mode

Mains power quality should be that of
a typical commercial or hospital
environment.

Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11

<5 % U

T

for 0.5 cycle

<5 % U

T

for 0.5 cycle

Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the HBP-
1300 requires continued operation
during power mains interruptions, it is
recommended that the HBP-1300 be
powered from an uninterruptible
power supply or a battery.

40 % U

T

for 5 cycles

40 % U

T

for 5 cycles

70 % U

T

for 25 cycles

70 % U

T

for 25 cycles

<5 % U

T

for 5 sec.

<5 % U

T

for 5 sec.

Power frequency
(50/ 60 Hz) magnetic
field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.

Note: U

T

is the a.c. mains voltage prior to application of the test level.