Omron Professional Blood Pressure Monitor HBP-1300 User Manual

Page 36

Advertising
background image

34

Immunity test

IEC60601-1-2

test level

IEC60601-1-2

test level

Electromagnetic

environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80

MHz

80% AM (2Hz)

3 Vrms

80 MHz to 2.5

GHz

80% AM (2Hz)

3 Vrms

3 V/m

Portable and mobile RF communications

equipment should be used no closer to any part

of the HBP-1300, including cables, than the

recommended separation distance calculated

from the equation applicable to the frequency of

the transmitter.

Recommend separation distance

d = 1.2
d = 1.2

80 MHz to 800 MHz

d = 2.3

800 MHz to 2.5 GHz

where P is the maximum output power rating

of the transmitter in watts (W) according to he

transmitter manufacturer and d is the recom-

mended separation distance in meters (m).

Field strengths from fixed RF transmitters as

determined by an electromagnetic site

survey

a

, should be less than the compliance

level in each frequency range

b

.

Interference may occur in the vicinity of

equipment marked with the following symbol:

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects, and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast

cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due

to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured

field strength in the location in which the HBP-1300 is used exceeds the applicable RF

compliance level above, the HBP-1300 should be observed to verify normal operation. If

abnormal performance is observed, additional measures may be necessary, such as reorienting

or relocating the HBP-1300.

b

Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.

P
P
P