Recommended Separation Distances:

Recommended separation distance between portable and mobile RF communications equipment
and the HBP-1300

The HBP-1300 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the HBP-1300 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the HBP-1300 as recommended below, according to the maximum output power of the communications
equipment.

Rated maximum output
power of transmitter
(W)

Separation distance according to frequency of transmitter
m

150 kHz to 80 MHz

d = 1.2

80 MHz to 800 MHz

d = 1.2

800 MHz to 2.5 GHz

d = 2.3

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.

Note1:At 80MHz and 800MHz, the separation distance for the higher frequency range applies
Note2:These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

P

P

P

HBP-1300.book Page 36 Wednesday, March 7, 2012 9:28 AM