3.1.13 Touch Screen Menu

The Main Menu screen gives you the
choice to use a ”Quick Set Up” (Fig
3.1.13a) - with settings you designate
for each procedure–or you may select
a pre-set program sorted in alphabeti-
cal order under the “Procedures”, or
under the “Specialty” categories, Fig.
3.1.13b. The settings are portrayed
with four adjustable settings: Mode,
Power, Time and Aiming Beam. The
settings can be changed as desired for
any of the program modes and saved
as“Favorites” if needed.

Mode
Mode refers to “continous wave” or “pulsed power,
Fig. 3.1.13c. Use the MODE key to toggle back and forth
between Continuous and Pulse as desired. The chosen mode
is displayed in gray color.

Power
Depressing the Power setting allows the operator to adjust the power set-
tings from 0 to 3 watts in 0.1W increments, Fig. 3.1.13d. This control is for
the working beam only. Hold the desired key to rapidly change the value.
Touching the right arrow increases the power, touching the left reduces the
power. Depress “Done” to return to the Main Menu.

Time
You have the ability to use the “Time” settings pre-set by procedure or modify the timer by using the arrow keys,
Fig. 3.1.13e. The laser will automatically shut off when the timer returns to 0. Touching the right arrow increases the
time, touching the left reduces the time. Depressing “Done” returns to the main menu.

Releasing the foot pedal in the middle of a
timed procedure will bring up a
“Procedure Interrupted!” window. The
timer will continue with the remaining
time if you press “Continue” or go back
to zero and start over if you press “Reset”,
Fig. 3.1.13f.