6.3.2 Recommended Autoclave Procedure

1.

Autoclave the handpiece sleeve only.
First remove the unit-dose tip after the procedure. Unscrew the metal
handpiece sleeve counterclockwise and separate from the handpiece.
The recommended autoclave cycle is 132°C at 27 psi for 15 minutes.
Fig 6.3.2a and b.

2.

Do not place the laser unit or the fiber cable in the autoclave!
The recommended procedure is to wipe the fiber cablewith a liquid
disinfectant (e.g. Lysol Spray Brand III). Please wear gloves. Do not
disconnect the fiber cable from the laser unit for cleaning procedure.
Fig 6.3.2c.

31

Navigator

3 W AT T D I O D E L A S E R

O

D

Y

S

S

E

Y

TM

®

SYSTEM COMPONENTS: PREPARATIONS, CARE AND MAINTENANCE

Section 6

Fig 6.3.2a

Fig 6.3.2b

Fig 6.3.2c