3.1.14 Preset Programs

The Odyssey Navigator comes with procedure specific settings installed.
The Procedures are accessed from the Main Menu by depressing
“Procedures.” Procedures may also be accessed through the “Specialty”
category. By accessing in this way, the procedures are sorted according to
speciality, Fig. 3.1.14.

After selecting a procedure, the settings may be manually adjusted to suit
your needs during a clinical procedure. The settings represent a good starting
point and can be adjusted at any time.

Program Screen Name

Tip Preparation

Technique

Mode

Power

Timer

Abscess

initiated

contact

Continuous

1.0 W

set to 0

Apthous Ulcer

un-initiated

non-contact

Pulse

1.4 W

30 seconds

Biopsy

initiated

contact

Continuous

1.0 W

set to 0

Contouring

initiated

contact

Continuous

0.8 W

set to 0

Denture Sore

un-initiated

non-contact

Pulse

2.0 W

30 seconds

Distal Wedge

initiated

contact

Continuous

1.5 W

set to 0

Draining of Abcesses

initiated

contact

Continuous

0.9 W

set to 0

Emergence Profile

initiated

contact

Continuous

0.8 W

set to 0

Expose Teeth

initiated

contact

Continuous

1.0 W

set to 0

Fibroma

initiated

contact

Continuous

1.0 W

set to 0

Flap

initiated

contact

Continuous

1.0 W

set to 0

Frenectomy

initiated

contact

Continuous

0.8 W

set to 0

Gingivectomy

initiated

contact

Continuous

1.0 W

set to 0

Gingivitis

initiated

contact

Continuous

0.4 W

20 seconds

Gingivitis Therapy

un-initiated

contact

Continuous

0.8 W

20 seconds

Gingivoplasty

initiated

contact

Continuous

1.0 W

set to 0

Hemostasis

un-initiated

non-contact

Continuous

1.0 W

set to 0

Herpetic Lesion

un-initiated

non-contact

Pulse

1.6 W

30 seconds

Hypertrophic tissue

initiated

contact

Continuous

1.5 W

set to 0

Implant exposure

initiated

contact

Continuous

1.0 W

set to 0

Lesion Removal

initiated

contact

Continuous

1.0 W

set to 0

Mucocele

initiated

contact

Continuous

0.9 W

set to 0

Operculectomy

initiated

contact

Continuous

1.5 W

set to 0

Ovate Pontic

initiated

contact

Continuous

1.0 W

set to 0

Periimplantitis

un-initiated

contact

Continuous

1.0 W

set to 0

Perio - Chronic

un-initiated

contact

Continuous

0.5 W

30 seconds

Perio - Severe

un-initiated

contact

Continuous

0.7 W

45 seconds

Socket Treatment

un-initiated

non-contact

Continuous

0.8 W

set to 0

Sulcular Debridement

un-initiated

contact

Continuous

0.8 W

30 seconds

Tissue Tag

initiated

contact

Continuous

1.5 W

set to 0

Tissue Welding

initiated

contact

Continuous

0.6 W

set to 0

Tongue Tie

initiated

contact

Continuous

1.0 W

set to 0

Troughing

initiated

contact

Continuous

0.8 W

set to 0

Vestibuloplasty

initiated

contact

Continuous

1.0 W

set to 0

18

T h e L e a d e r i n D i o d e L a s e r s .

LASER ASSEMBLY INSTRUCTIONS

Section 3

Fig 3.1.14

Advertising