5.4 Priming the Tubing

Pressing the PRIME key causes the tubing to fill with fluid.

1.

To prime the tubing, press and hold down the PRIME key. As long as you hold
down the PRIME key, fluid flows in the tubing, and the display shows
“Priming...”.

2.

To stop priming a tube, simply release the PRIME key. The pump motor stops
running, and fluid stops flowing.

Extensive testing has shown that the minimum motor speed required to prime
the tubing varies significantly with the size of the tubing. These variances are
shown in Table 9-3 in Chapter 9. If the tubing will not prime regardless of motor
speed, simply press on the pump stator (refer to Figure 3-11) while the stator is
shut and the motor is running. Pressing on the stator enhances the priming
action of the pump. Release pressure after the tubing is primed.

5.5 Calibrating the Tubing

Priming. . .

PRIME

ACTION

Pumps fluid at

current RPM or

FLOW RATE setting

as long as button is

pressed

Display shows

Operating the Pump Motor in the STOPPED Display Mode

5

5-6

Prime or START?

CALIBRATE

ACTION

Press PRIME to

prime tube or

START to dispense

desired amount of

fluid

Display shows

Priming...

PRIME

ACTION

Hold button down to dispense fluid

Display shows

Dispensing . . .

START

STOP

ACTION

Press Start/Stop button to

dispense fluid. Press again

to stop the process.

Display shows

ACTION

Scrolls up. Enter TOTAL VOLUME dispensed

xxx mL or xxx Liters

Display shows

CALIBRATED!

ACTION

Red LED next to

Calibrate key

lights up

CALIBRATE